Description

PW-Series High Quality Vertical Air-Circulating Drying Oven

 

Technical Specifications:

Name Vertical Air-Circulating Drying Oven
Model PW-9070A PW-9140A PW-9240A PW-9420A PW-9620A PW-9920A
Temperature Control Range Room Temperature +10~250℃
Constant Temperature Accuracy ±0.5℃
Fluctuation ±1℃;
Display Accuracy 1℃;
Heating Rate ≤5℃/min;(Can be arbitrarily adjusted to a speed lower than 5 degrees per minute)
Power Supply Voltage 220V/50HZ 380V/50HZ
Heating Power 1050W 1500W 2100W 4000W 4500W 6000W
Chamber Dimensions (Depth * Width * Height) 350*400*500 550*450*550 600*500*750 550*600*1300 600*800*1300 600*1000*1600
Overall Dimensions 610*600*755 770*600*845 815*650*1045 810*745*1670 860*945*1690 860*1140*2150
Packaging Dimensions 650*660*880 810*660*950 860*710*1150 900*850*1820 970*1100*1850 1000*1300*2250
Net Weight (KG) 48 67 77 115 145 190
Gross Weight (KG) 59 80 92 150 184 225
Shelf Configuration 2 pcs 3 pcs 4 pcs 5 pcs
Heating Tube Position back
Timing Range 1-9999 min
Safety Ventilation Hole Position rear
Operating Environment Temperature 5-40℃
Packaging Carton or Wooden Box
 

 

 

I. Application:

1.1 Widely used in the fields of biochemistry, agricultural research, environmental protection, chemical pharmaceuticals, medical and health, food, and machinery, it serves as a tool for drying glassware, conducting thermal denaturation and hardening experiments on samples, removing residual moisture from specimens, dry heat sterilization of glassware used for microbiology, preheating before heating experiments, experiments on hardening or softening food and chemical substances during heating, removing moisture to extend the shelf life of food under harsh conditions, and conducting drying, heat resistance tests on building materials and electronic components. It is also commonly used for drying and baking various materials in industrial and mining enterprises.

 

II. Performance Features: 

2.1 Mirror-like 304 stainless steel interior, rounded corners, with adjustable air inlets and outlets;

2.2 Maximum temperature of 250℃, with an accuracy of 1℃, and equipped with PID self-tuning function.

 

III. Chamber Structure and Materials:

3.1 Outer Chamber Material: The outer box is made of high-quality cold rolled steel treated with phosphoric acid film salt and then high-temperature spray plastic, with blue door and white body color.

3.2 Studio Material: Equipped with a mirror-like stainless steel interior, removable stainless steel electropolished shelves, tempered glass viewing window.

3.3 Internal Structure: Reasonable air duct structure and gentle airflow circulation design, with fan-forced heat dissipation circulation, ensuring good temperature uniformity in the working chamber, and a unique geometric cavity clean design.

3.4 Insulation Material: High-density glass fiber cotton is filled in the insulation layers on the front, back, left, right, top, and bottom of the box.

3.5 Sealing Method: The chamber door and the working chamber are sealed with high-temperature resistant silicon rubber seals to ensure the sealing of the working chamber. The inner panel of the chamber door is made of stainless steel. A double-layer tempered glass observation window is located on the door for observing changes in the samples inside the chamber. 3.6 Transparent and insulating, the chamber back has an air intake hole and a safety ventilation hole.

3.7 Temperature Port: The PT100 sensor enters from the back of the chamber.

3.8 Terminal Post: The wiring terminal for the heating elements is located at the bottom rear of the chamber.

3.9 Controller: Located at the top of the chamber, with an integrated control system.

3.10 Heating Element: Stainless steel heater.

 

IV. Temperature Control System and Features:

4.1 Custom-designed, digital display, microcomputer intelligent temperature controller with PID self-tuning function for high temperature control precision.

4.2 PT100 high-responsive thermal sensor and environmental scanning microprocessor chip form a high-precision, high-intelligence temperature control characteristic.

4.3 Built-in temperature calibrator to adjust minor temperature errors. Automatic alarm when working temperature deviates from the set temperature, with automatic power-off protection for heater malfunction.

4.4 Timer setting function, allowing any working time from 1 minute to 9999 minutes to be set, and the setting display window is closed at the end of the timing. (Temperature reaches the start time, time is up, automatic shutdown)

4.5 Both set parameters and actual parameters can clearly display set temperature, measured temperature, and timer time. Each parameter display is visible, with a modular design of the upper limit temperature tracking function. When the measured temperature deviates from the allowed range, the instrument will sound (light) alarm and automatically cut off heating.

 

V. Air Circulation System:

5.1 The circulation blower is located in the middle of the stainless steel heater on the back of the box,

5.2 Equipped with a long-shaft fan motor and multi-vane aluminum alloy blades to achieve horizontal convection and diffusion of air circulation within the box in an arc shape, ensuring uniform temperature throughout the box and maintaining stability.

 

VI. Heating System:

6.1 Stainless steel heaters are located at the back of the box, controlled by the PT100 sensor,

6.2 The heating output power of the heaters is uniformly controlled by a microcomputer within the intelligent instrument.

 

VII. Safety Protection System:

7.1 Fusible protection device

7.2 Chamber overheat protector, sensor fault alarm, upper limit overheat alarm

7.3 Fan overcurrent protection